The Effect of Intracochlear and Intratympanic Dexamethasone on Cochlear Implant Impedance
PDF
Cite
Share
Request
Original Investigation
P: 103-108
September 2023

The Effect of Intracochlear and Intratympanic Dexamethasone on Cochlear Implant Impedance

Turk Arch Otorhinolaryngol 2023;61(3):103-108
1. Department of Otolaryngology-Head and Neck Surgery, Pamukkale University Faculty of Medicine, Denizli, Turkey
2. Department of Biostatistics, Pamukkale University Faculty of Medicine, Denizli, Turkey
No information available.
No information available
Received Date: 17.06.2023
Accepted Date: 12.08.2023
Publish Date: 14.11.2023
PDF
Cite
Share
Request

ABSTRACT

Objective:

This study investigated the impact of different local corticosteroid applications on impedance measurements in patients with cochlear implants.

Methods:

The study was designed as a controlled, randomized, and prospective study in which 34 consecutive patients who had undergone cochlear implant surgery were divided into three groups. The first group received intracochlear dexamethasone, in the second group the middle ear cavity was filled with dexamethasone, and the third group did not receive dexamethasone. Intraoperative, postoperative 1st week, 1st month, 3rd month, 6th-month neural response telemetry, and impedances were measured. The measurements were compared by electrode groups representing the different regions of cochlea like basal (1–7), middle (8–13), and apical (14–22) regions.

Results:

The intergroup analysis showed no statistically significant differences in impedance measurements of the basal, middle, and apical regions (p>0.05). However, the impedances were lower in the two dexamethasone groups, especially in the basal and middle parts. Sixth month impedances were also lower in the dexamethasone groups. There was apparent stability in the impedance of the basal region with the intracochlear application during the first week.

Conclusion:

Local dexamethasone applications had a potentially positive impact on the impedance of the basal and middle regions. Patients had lower impedances than the control group during follow-up and at the endpoint. The increase in the apical region may indicate that dexamethasone was not reaching the apical zone in local applications.

Türkçe Özet

Koklear İmplant Empedansı Üzerinde İntrakoklear ve İntratimpanik Deksametazonun Etkisi

Amaç:

Bu çalışmada, koklear implant kullanan hastalarda farklı lokal kortikosteroid uygulamalarının empedans ölçümleri üzerindeki etkisi araştırıldı.

Yöntemler:

Kontrollü, randomize ve ileriye dönük olarak planlanan çalışmada koklear implant cerrahisi geçiren ardışık 34 hasta üç gruba ayrıldı. İlk grupta, kokleaya deksametazon uygulandı, ikinci grupta orta kulak boşluğu deksametazon ile dolduruldu, üçüncü grup ise deksametazon almadı. İntraoperatif, postoperatif 1. hafta, 1. ay, 3. ay, 6. ay NRT ve empedanslar ölçüldü. Ölçümler kokleanın bazal (1–7), orta (8–13) ve apikal (14–22) bölgeleri gibi farklı bölgelerini temsil eden elektrot gruplarına göre karşılaştırıldı.

Bulgular:

Gruplar arası analiz bazal, orta ve apikal bölgelerin impedansları arasında istatistiksel olarak anlamlı fark olmadığını gösterdi (p>0,05). Ancak, özellikle bazal ve orta kısımlarda deksametazon gruplarında empedanslar daha düşüktü. Altıncı aydaki empedanslar da deksametazon gruplarında daha düşüktü. İlk haftada intrakoklear uygulama ile bazal bölgede empedansta belirgin bir stabilite vardı.

Sonuç:

Lokal deksametazon uygulamalarının bazal ve orta bölgelerin empedansı üzerinde olası olumlu bir etkisi var. Hastalar takip sırasında ve sonlanım noktasında kontrol grubuna göre daha düşük empedansa sahipti. Apikal bölgedeki yüksek empedans artışı lokal uygulamalarda deksametazonun apikal bölgeye ulaşmadığı şeklinde yorumlanabilir.

Keywords:
Cochlear implants, dexamethasone, electrode impedance, fibrosis

Introduction

One of the common problems with a cochlear implant is the decrease in effectiveness over time due to physiological and mechanical damage in the cochlea (1). The short- and long-term effects of inflammation, osteoneogenesis, and fibrosis limited to the basal turn after surgery have been considered in this condition (2, 3). This is caused by a fibrotic capsule forming around the implant, which in turn results in an immune response to surgical trauma or a foreign body reaction to the platinum-iridium and silicon used in cochlear implants (4, 5). Studies have also shown a correlation between fibrotic tissue and electrode impedance (6). Choi et al. (7) suggested that impedance measurement could be used as a biomarker for cochlear damage.

Animal studies showed that glucocorticoids reduced foreign body reactions and fibrosis, thus increasing the lifespan of the spiral ganglia and the hair cells (8). Corticosteroids have been used for many years in cochlear implant surgeries. Although different systemic and local administration routes have been reported, it is still under development today (9, 10, 11).

We aimed to observe the effect of local corticosteroid applications during cochlear implant surgery on impedance.

Methods

A controlled, randomized, prospective clinical trial was planned in patients with cochlear implant surgery (registered with clinicaltrials.com, no: NCT04397354). Pamukkale University Non-invasive Clinical Research Ethics Committee approval was obtained (no: 60116787-020/20945, date: 23.03.2018). All patients signed the written informed consent form. Those with cochlear anomalies were excluded.

Cochlear implant operation was performed under general anesthesia, and 1 mg/kg methylprednisolone was administered intravenously to all patients in addition to anesthetic drugs as a part of routine general anesthesia. The round window soft technique was used to implement the devices. The cases with cochleostomy were excluded. The same cochlear implant electrode model was used in all patients (Cochlear, Inc. CI 422). The cochlear implants were activated after one month.

The patients were randomly divided into three groups for implantation using the random number table method according to the administration of dexamethasone (dex) (4 mg/mL).

1. In the first group, dex was administered slowly with a 27-gauge needle into the cochlea after a round window membrane incision (Group 1, the Cocdex group).

2. In the second group, dex was administered into the middle ear after a round window membrane incision (Group 2, the Middex group).

3. Dex was not administered to the third group (Group 3, control, the Nodex group).

The drug was left in place for three minutes before inserting the electrodes.

Intraoperative neural response telemetry thresholds and impedances were measured. Impedance measurements were repeated at the end of the first postoperative week, and at the first third, and sixth months. Monopolar1+2 (MP1+2)  impedance (kOhm), measurements were used for comparison.

The mean of the basal (1–7), middle (8–13), and apical (14–22) electrodes were used for comparison. We also compared the average of all electrodes. The Custom Sound EP 5.0 (5.0.4.136) program provided by Cochlear, Inc. was used for measurement.

Results

Initially, a total of 34 patients were included in the study. Of these, 12 were in Group 1, 12 were in Group 2, and 10 were in Group 3. One patient in Group 1 and one patient in Group 3 were excluded from the study due to noncompliance with follow-up and repetitive measurement times. Three patients in Group 3 in whom the cochleostomy technique was used were also excluded. Due to the pandemic, the 1st-week measurement of one patient, the 1st-month measurement of one patient, the 3rd-month measurement of two patients in Group 1; the 1st-month measurement of one patient, the 3rd-month measurement of one patient in Group 2; and the 1st-week measurements of two patients, the 1st-month measurement of one patient in Group 3 could not be performed on the planned date, the values of these measurements were not used in the analysis (Figure 1).

Figure 1

There were no differences between the groups regarding gender (17 women and 12 men, p=0.543) and age (p=0.688).

The variations of the mean MP1+2 impedance measurements over time are given in Figure 2.

Figure 2

In the apical zone, impedances reached the highest level, usually at the end of the first month, except in the Cocdex group, and then decreased rapidly in all groups.

In the basal and middle zones, all three groups reached the maximum impedance value at the end of the first month. In the first week, the impedances measured from the basal region were lowest in the intracochlear group (Figure 2). In the first month, impedances were higher in the Nodex group compared to the two dex groups in the basal and middle regions (Figure 2). The statistical comparisons of the time points within the groups are also plotted in Figure 2. Impedances increased until the first month (between t0 and t2) in all regions except for the basal and middle electrodes of the Cocdex roup (p>0.05), and these increases were statistically significant (p<0.05).

Figure 2

No statistically significant results were found between the groups at any time point or in any region (p>0.05) (Table 1). The change in the apical region in the 1st week was not significantly faster than in the middle and basal parts. The impedances were lower in the first week in the middle and basal regions. Their highest values were measured at the end of the first month in all three groups. Impedances were higher in the Control (Nodex) group but were not statistically significant (p>0.05).

Table 1

There were also no statistically significant differences when the mean impedances of all electrodes were compared (p>0.05).

Discussion

In this prospective, randomized, controlled study, we aimed to measure the effect of different dex applications on tissue inflammation during cochlear implantation by electrode impedance. We found that the intraoperatively measured impedances decreased in the long-term and reached the lowest point, mainly in the 6th month, which was the study time limit. Impedances were lower in the dex groups than in the control group in the basal and middle regions. The stability of impedance at basal electrodes in the first week of the Cocdex group was marked. The first-week impedances of the Middex group were also lower than those of the Nodex group. These findings may be attributed to the local and short-duration effects of corticosteroids. The sharp increase in impedance in the apical electrodes during the first week, even in the dex groups, may indicate that dex was not reaching the apical zones. In all groups, a marked decrease in impedances was observed after the first month, the cochlear implant’s activation date. Again, the impedances of the dex groups decreased more than the those of the control (Nodex) group at the endpoint of the study.

A similar hypothesis was tested in guinea pigs. In this study the authors compared the effects of the intratympanic, intracochlear, and systemic administration routes by cytokines and residual hearing. They reported that the intracochlear route had reached the highest drug concentration. Intracochlear dex provided better protection for residual hearing and a less inflammatory response in the cochlea (10). We observed some promising results, especially in the basal region.

The results of animal and human studies with local steroid applications elicited the research on dex eluting electrodes. Astolfi et al. (12) tried a 10% dex eluting electrode on guinea pigs and reported that less tissue growth had been observed. Briggs et al. (3) implanted a dex eluting electrode and followed the patients for two years with MP1+2 impedance measurements. They found that the experimental electrode had lower impedance at all time points and all cochlea regions than the standard electrode. They observed a direct reduction of impedances even in the first week. Our study also found stability in the impedances in the basal region. After the effect of dex had diminished over time, the impedances rose again.

Additional systemic steroids used with dex eluding electrodes were tested (13). Systemic steroids did not show an additional effect over inserting electrodes alone. The authors concluded that the protective effect of steroids was prominent, especially in traumatic insertions. We could not obtain statistically significant results for the use of dex. Perhaps this was due to the atraumatic insertion technique in all patients. Ahmadi et al. (14) also support this conclusion. They said there must be trauma in the cochlea to see the effectiveness of dex. Non-traumatic insertion preserved the cochlea in most animals.

Lee et al. (2) compared different local and systemic applications in an animal implant model. They reported no significant differences between the delivery routes, but that could be effective if used for a longer duration and higher dose. The protective effect of dex had a linear relationship with the concentration of the drug and the time of contact with the round window. Chang et al. (9) reported that 2% dex for 60 minutes had the same protective effect with 20% dex for 30 minutes on guinea pigs.

We observed an increase in impedance in the apical region in the early period in both dex groups. Wei et al. (15) also observed the same increase in the 1st week with an early switch-on technique. In the 8th week, they showed that there was an increasing trend of impedance in all parts of the cochlea and was highest in the basal region. They argued that this was because fibrosis had started on days 2 to 5 after the operation, fibrotic tissue began to dissolve in the second week, and more severe fibrotic tissue formed in the basal region due to trauma. In our study, impedance remained low in the dex groups in the basal region. It might be due to the anti-inflammatory effect of the corticosteroids.

One of the limitations of this study was the coronavirus disease-2019 pandemic. Some patients missed the exact test dates, so we omitted some data from the calculations. Another limitation is the short time between dex administration and implant placement.

Nevertheless, this kind of study in patients is rare. Its advantages are that standard commercial electrodes were used, all operations were done by a single surgeon and the tests by a single audiologist, and only objective parameters were measured.

Conclusion

Although there were no significant differences among the groups, intracochlear and intratympanic applications of dex positively impacted the impedance in the basal and middle regions during the first week. Patients in the dex groups had lower impedances than the control group during follow-up and at the endpoint. The sharp increase in the apical region may indicate that dex was not reaching the apical zone in local applications.

Main Points

• Local dexamethasone applications had a possible positive impact on the impedance of the basal and middle regions.

• Patients in the dexamethasone groups had lower impedances than the control group during follow-up and at the endpoint.

• The impedance increase in the apical region may indicate that dexamethasone was not reaching the apical zone in local applications.

Statistical Analysis

All statistical analyses were performed with the SPSS 25.0 software [IBM SPSS Statistics 25 software (IBM Corp.: Armonk, NY, USA)]. Continuous variables were expressed as mean ± standard deviation, median (minimum-maximum values), and categorical variables as number and percent. The Shapiro–Wilk test was used to test for normality. If parametric test conditions were satisfied, the One-Way Analysis of Variance (post-hoc: Tukey test) was used to compare groups. If parametric test conditions were not satisfied, Kruskal–Wallis variation analysis (post-hoc: The Mann–Whitney U test with Bonferroni correction) was used to compare the groups. For pairwise comparisons, if parametric test conditions were satisfied the Repeated Measures ANOVA (post-hoc: Bonferroni test), and if parametric test conditions were not satisfied, the Friedman (post-hoc: Wilcoxon signed-rank test with Bonferroni correction) tests were used. The chi-square test was used to compare categorical variables and p<0.05 was considered statistically significant.

References

1
Scheper V, Hessler R, Hütten M, Wilk M, Jolly C, Lenarz T, et al. Local inner ear application of dexamethasone in cochlear implant models is safe for auditory neurons and increases the neuroprotective effect of chronic electrical stimulation. PLoS One 2017; 12: e0183820.
2
Lee MY, Kim YC, Jang J, Jung JY, Choi H, Jang JH, et al. Dexamethasone delivery for hearing preservation in animal cochlear implant model: continuity, long-term release, and fast release rate. Acta Otolaryngol 2020; 140: 713-22.
3
Briggs R, O 'Leary S, Birman C, Plant K, English R, Dawson P, et al. Comparison of electrode impedance measures between a dexamethasone-eluting and standard Cochlear™ Contour Advance® electrode in adult cochlear implant recipients. Hear Res 2020; 390: 107924.
4
Cho HS, Lee KY, Choi H, Jang JH, Lee SH. Dexamethasone is one of the factors minimizing the inner ear damage from electrode insertion in cochlear implantation. Audiol Neurootol 2016; 21: 178-86.
5
Wilk M, Hessler R, Mugridge K, Jolly C, Fehr M, Lenarz T, et al. ımpedance changes and fibrous tissue growth after cochlear implantation are correlated and can be reduced using a dexamethasone eluting electrode. PLoS One 2016; 11: e0147552.
6
Needham K, Stathopoulos D, Newbold C, Leavens J, Risi F, Manouchehri S, et al. Electrode impedance changes after implantation of a dexamethasone-eluting intracochlear array. Cochlear Implants Int 2020; 21: 98-109.
7
Choi J, Payne MR, Campbell LJ, Bester CW, Newbold C, Eastwood H, et al. Electrode impedance fluctuations as a biomarker for inner ear pathology after cochlear implantation. Otol Neurotol 2017; 38: 1433-9.
8
Kather M, Koitzsch S, Breit B, Plontke S, Kammerer B, Liebau A. Metabolic reprogramming of inner ear cell line heı-oc1 after dexamethasone application. Metabolomics 2021; 17: 52.
9
Chang A, Eastwood H, Sly D, James D, Richardson R, O'Leary S. Factors influencing the efficacy of round window dexamethasone protection of residual hearing post-cochlear implant surgery. Hear Res 2009; 255: 67-72.
10
Lyu AR, Kim DH, Lee SH, Shin DS, Shin SA, Park YH. Effects of dexamethasone on intracochlear inflammation and residual hearing after cochleostomy: a comparison of administration routes. PLoS One 2018; 13: e0195230.
11
Eastwood H, Chang A, Kel G, Sly D, Richardson R, O'Leary SJ. Round window delivery of dexamethasone ameliorates local and remote hearing loss produced by cochlear implantation into the second turn of the guinea pig cochlea. Hear Res 2010; 265: 25-9.
12
Astolfi L, Simoni E, Giarbini N, Giordano P, Pannella M, Hatzopoulos S, et al. Cochlear implant and inflammation reaction: safety study of a new steroid-eluting electrode. Hear Res 2016; 336: 44-52.
13
Chambers S, Newbold C, Stathopoulos D, Needham K, Miller C, Risi F, et al. Protecting against electrode insertion trauma using dexamethasone. Cochlear Implants Int 2019; 20: 1-11.
14
Ahmadi N, Gausterer JC, Honeder C, Mötz M, Schöpper H, Zhu C, et al. Long-term effects and potential limits of intratympanic dexamethasone-loaded hydrogels combined with dexamethasone-eluting cochlear electrodes in a low-insertion trauma guinea pig model. Hear Res 2019; 384: 107825.
15
Wei JJ, Tung TH, Li LP. Evolution of impedance values in cochlear implant patients after early switch-on. PLoS One 2021; 16: e0246545.